Spartan RX

Technology

Workflow

Workflow: Cheek swab

Sample Type: <4 minutes

Hands-on Time: None

Sample Preparation: None

Run Time: <60 minutes

Method Comparison

A total of 325 individuals were tested with the Spartan RX CYP2C19 System and compared with bi-directional DNA sequencing.

If the first test result (First Pass) was a “No Call”, the test was repeated (Second Pass).

Reproducibility

The Spartan RX CYP2C19 System was evaluated at three sites. At each site, operators repeatedly tested eight individuals with the following genotypes: *1/*17, *2/*3, *1/*1, *17/*17, *1/*3, *2/*17, *2/*2, and *1/*2. Genotypes were confirmed with bi-directional DNA sequencing.

If the first test result (First Pass) was a “No Call”, the test was repeated (Second Pass).

Number of Samples Tested: 960

First Pass Call Rate: 98.8%

Second Pass Call Rate: 99.8%